This online hydrogen-in-oxygen analyzer features high sensitivity, simple operation, and excellent sensor stability. It is less affected by temperature, flow rate, and pressure, and is not influenced by factors such as vibration and installation angle. It has low requirements for the content of impurities like moisture and dust in the sample gas, thus enjoying wide applications.
-
Technical Specifications
Sensor: Constant temperature sensor with temperature compensation
Measurement Range: 0~99.99% or customized
Typical Accuracy: ±0.01%
Repeatability: ±0.01%
Sample Gas Flow: 200~300mL/min
Response Speed: 10~15 seconds
Operating Temperature: -20℃~50℃
Operating Pressure: ≤1.0MPa
Signal Output: 4~20mA (0~2% or customized), ±0.05mA 300Ω
Display: LCD remote display
Alarm Output: 2 channels of alarm dry contact output. Factory default settings: Alarm 1 (high alarm, alarm value 1.5%), Alarm 2 (low alarm, alarm value 0.2%)
Power Supply: DC24V, power consumption ≤3W
Inlet/Outlet Connection: Φ6, stainless steel pipe